Chapter 22:Technical Specifications
User Manual
22-10
Gas or vapour
Gas level
CO2
Agents
N2O
ISA CO2
ISA
AX+/OR+
N2O-note4)
60 vol%
- note2)
- note1)
- note1)
- note1)
HAL-note4)
4 vol%
- note1)
- note1)
- note1)
- note1)
ENF, ISO, SEV-note4)
5 vol%
+8% of reading-note3)
- note1)
- note1)
- note1)
DES-note4)
15 vol%
+12% of reading-note 3)
- note1)
- note1)
- note1)
Xe (Xenon)-note4)
80 vol%
-10% of reading-note3)
- note1)
- note1)
He (Helium)-note4)
50 vol%
-6% of reading-note3)
- note1)
- note1)
Metered does inhaler propellants-note4)
Not for use with metered dose inhaler propellants
C2H5OH (Ethanol)-note4)
0.3 vol%
- note1)
- note1)
- note1)
- note1)
C3H7OH (Isopropanol)-note4)
0.5 vol%
- note1)
- note1)
- note1)
- note1)
CH3COCH3 (Acetone)-note4)
1 vol%
- note1)
- note1)
- note1)
- note1)
CH4 (Methane)-note4)
3 vol%
- note1)
- note1)
- note1)
- note1)
CO (Carbon monoxide)-note5)
1 vol%
- note1)
- note1)
- note1)
- note1)
NO (Nitrogen monoxide)-note5)
0.02 vol%
- note1)
- note1)
- note1)
- note1)
O2-note5)
100 vol%
- note
٢
)
- note2)
- note1)
- note1)
Note 1 : Negligible interference, effect included in the specification “ Accuracy all conditions” above.
Note 2 : Negligible interference with N2O/O2concentrations correctly set, effect included in the specification “ Accuracy all
conditions” above.
Note 3 : Interference at indicated gas level. For example, 50 vol% Helium typically decreases the CO2 readings by 6%. This
means that if measuring on a mixture contaning 5.0 vol% CO2and 50 vol% Helium, the actual measured CO2 concentration
will typically be (1-0.06) * 5.0 vol% = 4.7 vol% CO2.
Note 4 : According to the EN ISO 80601-2-55:2011 standard.
Note 5 : In addition to the EN ISO 80601-2-55:2011 standard.
BFA (Brain Function Assesment)
BFA Interface
Required for Integratig BFA mdule and monitors
EEG sensitivity
±450µV
Noise
<2µVp-p <0.4µV RMS, 0.25-250 Hz
CMRR
>140dB
Input impedance
>50M
Ω
Sample rate
1000 samples/sec(16 bits equivalent)
Brain Function Index (BFI)
0-100. Filter 1-47Hz, 1sec. update
EMG
0-100. Filter 30-47 Hz,1 sec. update
BSR
0-100. Filter 2-47 Hz, 1 sec. update
Signal Quality Index (SQI)
0-100. 1 sec. update
EEG Waveform
±250µV, user-adjustable, 5 sec
Alarms
Auditory and visual, user-adjustable limits
Artifact rejection
Automatic
Sensor impedance measurement
0-30kOhm / Manual-Automatic/ measurement current 0.06µ A
Power supply
5 VDC
Power Consumption
Less than 0.5 W
Weight
100 gr
Dimensions
111×64×25 mm
Classification
Class I, type BF, continuous use
Sensors
Ambu Neuro Sensors
Cable length
195 cm/ 77” with 35 cm/ 14” split
Memory
Data recording (96 hours)
Trend
BFI/EMG/SQI/BS, 10 sec. update
Environment - Operation
Temperature
5-40°C