Chapter 14: GAS Monitoring (Sidestream)
User Manual
14-13
The Capnography Window for Sidestream sensor in different modes is as follows:
CO2 WINDOW in CO2(ONLY) mode
GAS WINDOW in AX+ mode
GAS WINDOW in OR+ mode
Capnography Window of sidestream sensor in different modes
■
Co2 UNIT
Pick this item to adjust CO2 measurement unit. (Options: mmHg, KPa, %V)
EtCo2 in %V is the EtCo2 value (in mmHg) divided by ambient barometric pressure
(in mmHg) which is a percentage of the barometric pressure.
EtCo2(%V) =
)
(
)
(
2
mmHg
Brometric
mmHg
EtCo
p
P
EtCo2(KPa) =
1000
322
.
133
)
(
2 mmHg
EtCo
P