Chapter 9: SpO2 & Rainbow Parameters Monitoring
User Manual
9-24
excessive motion or other signal
interference.
application.
3-Change the sensor
site.
background becomes gray and the
alarm is disabled and ignores this
fault.
SPO2 LOW
SPMET
CONFIDENCE
SpMet measurement does not
have confidence due to poor
signal quality caused by
excessive motion or other signal
interference.
1-Assess the patient.
2-Check the sensor and
ensure proper sensor
application.
3-Change the sensor
site.
Alarm level 3- the message is
displayed in cyan background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
SPO2 LOW
SPHB
CONFIDENCE
SpHb measurement does not
have confidence due to poor
signal quality caused by
excessive motion or other signal
interference.
1-Assess the patient.
2-Check the sensor and
ensure proper sensor
application.
3-Change the sensor
site.
Alarm level 3- the message is
displayed in cyan background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
After taking the mentioned actions if above messages are displayed again, the SpO2 probe may
be damaged and you should contact local After Sales Services.
c) Messages
SPO2 MESSAGES
Message
Cause
Solution
Explanation
SPO2 CABLE
NEAR EXP
The SpO2 cable is near
expiration
.
In this condition SPO2 parameter is
displayed.
SPO2 SENSOR
NEAR EXP
The SpO2 sensor is near
expiration
.
In this condition SPO2 parameter is
displayed.
SPO2 AD
SENSOR NEAR
EXP
The SpO2 adhesive sensor is
near expiration
.
In this condition SPO2 parameter is
displayed.