Chapter 14: GAS Monitoring (Sidestream)
User Manual
14-16
NOTE:
For enabling ISA sensor, you can enter Gas window and set the monitor to
Measure mode.
■
GAS/RESP
Select this item to determine that respiration evaluation is performed by "Multi-gas"
or "RESP" module. Available options are "GAS" and "RESP". When selecting
"RESP", the system switches Multi-gas module to standby mode, and displays RESP
waveform and parameters.
■
FILL SIGNAL
Pick "ON" to show the waveform in filled form.
■
CO2 ALARM
Pick "CO2 ALARM" in GAS WINDOW to call up the following menu:
CO2 ALARM
Pick "ON" to enable alarm functions such as parameters blinking, audio alarm and
light indicator. Pick "OFF" to disable the alarm functions and there will be a "
"
symbol in Multi-gas Parameter Area.
ALARM LEVEL
Selectable between 1 and 2. Level 1 represents the most serious case.
EtCO2 LIMIT
Alarm is activated when the EtCo2 exceeds adjusted ALARM HIGH or LOW limit
(Range: 0.4~13%V step 0.1%V)
Default for upper limit is 6.5%V and for lower limit is 2.6%V.
FiCO2 HIGH