Chapter 24: Care and Cleaning
User Manual
24-8
Preventive Maintenance (PM) Checklist
SAADAT Co.
Form No. : PL-F-24
PM Form (BEDSIDE)
State: City: Medical center:
Ward:
Device model: Serial number:
Installation date: Inspection date:
N/A
NOK
OK
Test and Inspection Item
No.
No damage or breakage in the back case and panel
Visual inspection
1
Cleaning and disinfection according to the user manual
Correct function
Rotary
2
Correct function
Keyboard
3
Correct function
Touch
4
Correct display of Waveform area, Parameter area and
Message area
Display screen
5
Unplugging the system (check the battery function)
Battery
6
Check the fuse
Periodic usage of the battery
Alarm activation
Alarm
7
Clarity of alarm sound
Correct function of alarm LEDs
Saving date& time settings
Setup
8
Check ECG cable (clamps, leadwire, trunk)
ECG
9
Check ECG window (Pacemaker, beat sound, etc)
Cleaning and disinfection according to the user manual