Chapter 10: NIBP Monitoring
User Manual
10-9
■
NIBP UNIT
Pick this item to adjust measurement unit. (Options: mmHg or KPa)
■
ALARM LEVEL
Selectable between 1 and 2. Level 1 represents the most serious case.
■
NIBP ALARM
Pick "ON" to enable NIBP alarm functions such as parameters blinking, audio alarm and
light indicator. Pick "OFF" to disable the alarm functions and there will be a "
"symbol
in the Parameter Area.
■
NIBP SYS LIMIT
SYS alarm is activated when the systolic pressure exceeds adjusted ALARM HIGH limit
or falls below adjusted ALARM LOW limit.
Adult
→
Max: 255, Min: 30
Pediatric
→
Max: 240, Min: 30
Neonate
→
Max: 135, Min: 30
■
NIBP DIA LIMIT
DIA alarm is activated when the diastolic pressure exceeds adjusted ALARM HIGH limit
or falls below adjusted ALARM LOW limit.
Adult
→
Max: 220, Min: 15
Pediatric
→
Max: 220, Min: 15
Neonate
→
Max: 110, Min: 15
■
NIBP MAP LIMIT
MAP alarm is activated when the mean arterial pressure exceeds adjusted ALARM
HIGH limit or falls below adjusted ALARM LOW limit.(min:25 and max:200)
Adult
→
Max: 235, Min: 20
Pediatric
→
Max: 230, Min: 20
Neonate
→
Max: 125, Min: 20
■
AUTO/MANUAL/STAT
There are three modes of measurement available: MANUAL, AUTO and STAT. In the
MANUAL mode, only one measurement is performed. In the AUTO mode, measurement
is repeated over a specified period of time ; available intervals
are
1,2,3,5,10,15,20,30,45, 60, 90 minutes and 2, 4, 8, 12,16,20,24 hours. In STAT mode,