Chapter 22:Technical Specifications
User Manual
22-3
Reject from heart rate counter
Re-insert into ECG to display on screen
Ineffective pace rejection
HR:0, Pace: 60
HR:60, Pace:60
HR:30, Pace:80
Beside rejection of atrial paces preceed ventricular paces by 150 or 250 ms
Protection
Defibrillator and Electrosurgery
ANALOG OUTPUT
Signals
ECG
Maximum delay
≤30 ms
Output range
± 5 V
Signal gain
1000 (1V/mV)
Gain accuracy
± 20 mV
Maximum offset
± 50 mV
ECG bandwidth
“MONITOR”
0.5 - 24 Hz
“NORMAL”
0.5 - 40 Hz
“EXTENDED”
0.05-100 Hz
Pacemaker pulses
Amplitude:
5 V (nominal)
Duration:
5 ms
ECG range
-5 to 5 mV
Output impedance
249 Ω ± 5%
Data rate
400 samples/sec
ARRHYTHMIA ANALYSIS
Type
ASYS, VFIB, VTAC, RUN, AIVR, COUPLET, BIGEMINY, TRIGEMINY,
TACHY, BRADY, AFIB, PAUS, FREQUENT PVCs
Learning
Rapid Learning: only 20 seconds required for recognition of dominant
rhythm.
Method
Real time arrhythmia detection with innovative feature.
Memory
Capability of storing the latest 150 ARR event (waveform and Parameters)
ST ANALYSIS
Display resolution
0.01 mV
Measurement Range
-2mv to +2mv
Alarm Range
-2mv to +2mv
Features
User Adjustable Isoelectric and ST point trending of ST values
Update period
5 Sec.
NIBP
Measurement method
Oscillometric
Measurement mode
Manual/Automatic/Stat
Measurement time
20-25 sec (excluding cuff inflation time)
Cuff pressure rang
Adult
0-290 mmHg
Neonate
0-145 mmHg
Measurement Range
Adult
SYS
30 ~ 255 mmHg
DIA
15 ~ 220 mmHg
MAP
20 ~ 235 mmHg
Neonate
SYS
30 ~ 135 mmHg