Chapter 22:Technical Specifications
User Manual
22-11
Rel humidity
20~96%
Altitude
-200~3000m
Cardiac Output
Method
Right Heart Thermodilution
Range
0.5-18 l/min
Resolution
0.01l/min
Reproducibility
±3%
Recorder
Model
Internal Thermal Recorder SP58
Channel
Up to 3 waveforms
Printing Speed
6,12.5,25 mm/sec
Paper Size
57mm by 59 foot roll
DRUG CALCULATE
To calculate the dose and time of medication
ALARM
Sources
Error messages, All other parameter limits
Alarm On/Off
Selectable for all parameters
Alert
Blinking on Display, Volume Selectable Audio Alarms, Light indicator
ALARM RECALL
Displaying occurred alarms along with ECG/SpO2/2IBP/RESP/GAS waveforms (20 recent alarm)
BED TO BED
“Bed to Bed” function
is intended to display information of one bedside connected to the central system on target bedside.
TREND
Sources
HR,PVCs,ST,AFIB,SpO2, RR, T1, T2, IBP1(SYS,DIA,MAP),
IBP2(SYS,DIA,MAP), IBP3(SYS,DIA,MAP), IBP4(SYS,DIA,MAP),
EtCo2,FiCo2,AWRR(sidestream, mainstream),
EtN20,FiN2O,EtO2,FiO2,EtAA,FiAA(ISO, DES, ENF, HAL, SEV)
Trend Time Save
96 Hours
Trend Time Interval
15, 30, 45 Min, 1, 2 and 4 Hours
Resolution
1 sec
OXY-CRG
6 Parameters Trend
INPUT/OUTPUT
Network
TCP/IP Protocol Ethernet LAN with RJ45 Interface
VGA Connection
VGA output with same page