Chapter 9: SpO2 & Rainbow Parameters Monitoring
User Manual
9-20
SpMet LOW
SpMet violates adjusted low
alarm limit
● SpMet value blinks.
●
The alarm indicator flashes
●
The alarm message is
displayed in a background
corresponding to its level.
Activated
SpHb HIGH
SpHb violates adjusted high
alarm limit
● SpHb value blinks.
●
The alarm indicator flashes
●
The alarm message is
displayed in a background
corresponding to its level.
Activated
SpHb LOW
SpHb violates adjusted low
alarm limit
● SpHb value blinks.
●
The alarm indicator flashes
●
The alarm message is
displayed in a background
corresponding to its level.
Activated
b) Technical alarms
SPO2 ALARMS
Alarm
Cause
Solution
Explanation
SPO2 NO
CABLE
SpO2 cable is not fully inserted
to the patient monitor system.
Make sure that the
SpO2 cable is correctly
connected into the
monitor.
Alarm level 3- the message is
displayed in cyan background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
SPO2
REPLACE
CABLE
The life of the SpO2 cable has
expired.
Replace the SpO2
cable.
Alarm level 2- the message is
displayed in yellow background.
By pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
SPO2 CABLE
DEFECT
1. The SpO2 cable is damaged
2. SpO2 cable is not compatible.
1. Make sure that the
Masimo SpO2 cable is
correctly connected
into the monitor.
2. Restore power to the
instrument. If this
Alarm level 2- the message is
displayed in yellow background.
By pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.