2. Description of the Piranha
Hardware and Specifications
2014-06/5.5C
19
Piranha & QABrowser Reference Manual
Computed Tomography
Parameter
RQ
Range
Inaccuracy
Resolution
kVp (standard)
W / 3.0 mm Al
W / 3 mm Al + 1.2 mm Ti
(Siemens Straton (Siem1))
2
GECT (7°)
2, 3
Acquillion 64
(Toshiba)
2
GECT (10.5°)
2, 3
C1
C3
2
C4
2
C5
2
C6
2
45 – 160 kV
75 – 145 kV
75 – 145 kV
75 – 145 kV
65 – 150 kV
±1.5 %
4 digits
(10 or 100 V)
Irradiation time
0.1 ms – 2000 s
1 – 65535 pulses
±1 % or ±0.5 ms
±1 pulse
0.5 ms
1 pulse
Air kerma (Dose)
-
-
-
Estimated total
filtration
C1
C6
1.5 – 90 mm Al
(75 – 160 kV)
±10 % or ±0.3 mm
±15 % >50 mm Al
(75 – 120 kV, HF/DC,
>10 µGy/s)
2 digits
(0.1 or 1 mm)
Quick-HVL
C1
1.2 – 14 mm Al
(75 – 150 kV)
±10 % or ±0.2 mm
(75 – 120 kV, HF/DC,
>10 µGy/s)
1
3 digits
(0.01 or 0.1 mm)
Note 1
: This is valid for a tube with 14° anode angle. The HVL for a 22° anode is typically 0,5 mm
lower (@ 80 kV, 3 mm TF).
Note 2
: The C3 and higher numbered calibrations are only available for product versions 2.0 or
higher.
Note 3
: The C4 (
GECT (7°)
) is suitable for all GE CT tubes which have a 7° anode angle as well as
other manufacturers CT tubes and replacement tubes with a 7° anode angle. The C6 (
GECT
(10.5°)
) is suitable for GE CT tubes with a 10.5° anode angle..
Pulses
Summary of Contents for Piranha
Page 1: ...Reference Manual English Version 5 5C Piranha QABrowser RTI article number 9629050 10 ...
Page 8: ...Introduction Chapter 1 ...
Page 12: ...Description of the Piranha Chapter 2 ...
Page 40: ...Description of the QABrowser Chapter 3 ...
Page 77: ...Measurement Principles Theory Chapter 4 ...
Page 91: ...Measurements with the Piranha System Chapter 5 ...
Page 151: ...Optional Accessories Chapter 6 ...
Page 155: ...Problems and Solutions Chapter 7 ...
Page 161: ...Glossary Chapter 8 ...
Page 181: ...Notes Piranha QABrowser Reference Manual 2014 06 5 5C Notes 177 ...