5. Measurements with the Piranha System
Dental and Panoramic Dental
Piranha & QABrowser Reference Manual
2014-06/5.5C
130
The Piranha now first analyses the beam and displays the type of waveform. This is
done once. Then the estimated total filtration is displayed.
Read the measured values from the display.
6. Repeat the measurement for other generator settings or select an application to
measure further.
5.6.2
Waveforms
The example below explains why a change of delay change the value of measured kVp
and the kV and radiation waveform on a one phase dental unit.
Delay = 0 ms
Delay = 200 ms
Delay = 500 ms
From the study of the above three screens several conclusions can be made:
1. A stable output level is not reached until after approximately 200 to 300 ms (20 to 30
pulses for a 50 Hz main based dental unit).
2. The exposure time is depending on the definition of the trig level.
3. The kVp value in the RTD is related to selected delay and window and is several kV
higher in the beginning of the exposure at the same time as the radiation level is
relative low here. Therefore a delay of 200 ms is default for the Piranha.
If the signal to the detector is too low to give a correct kV value, this part of
the waveform will be blank. This is the reason why only the upper parts of
Summary of Contents for Piranha
Page 1: ...Reference Manual English Version 5 5C Piranha QABrowser RTI article number 9629050 10 ...
Page 8: ...Introduction Chapter 1 ...
Page 12: ...Description of the Piranha Chapter 2 ...
Page 40: ...Description of the QABrowser Chapter 3 ...
Page 77: ...Measurement Principles Theory Chapter 4 ...
Page 91: ...Measurements with the Piranha System Chapter 5 ...
Page 151: ...Optional Accessories Chapter 6 ...
Page 155: ...Problems and Solutions Chapter 7 ...
Page 161: ...Glossary Chapter 8 ...
Page 181: ...Notes Piranha QABrowser Reference Manual 2014 06 5 5C Notes 177 ...