
5. Measurements with the Piranha System
Dental and Panoramic Dental
2014-06/5.5C
129
Piranha & QABrowser Reference Manual
2. Place the detector on the table at
the distance that is clinically
relevant. As default, the kV-filter for
55 - 105 kV is selected
automatically. Furthermore the
Piranha surface should optimally be
placed perpendicular to the focal
spot, see also
Angular Sensitivity,
Piranha
.
3. It is recommended to make a check measurement to confirm that the detector area is
uniformed radiated. Select a exposure time of at least 400 ms. The Piranha
automatically changes back to the previous selected kV-filter after the check
exposure. As default, this is radiography filter R2 indicated by a [4] on the
QABrowser screen.
4. Set the kV and mA/time (or mAs) to desired values. Select a exposure time to near
400 ms since the delay is set to 200 ms.
5. Make an exposure.
24
Summary of Contents for Piranha
Page 1: ...Reference Manual English Version 5 5C Piranha QABrowser RTI article number 9629050 10 ...
Page 8: ...Introduction Chapter 1 ...
Page 12: ...Description of the Piranha Chapter 2 ...
Page 40: ...Description of the QABrowser Chapter 3 ...
Page 77: ...Measurement Principles Theory Chapter 4 ...
Page 91: ...Measurements with the Piranha System Chapter 5 ...
Page 151: ...Optional Accessories Chapter 6 ...
Page 155: ...Problems and Solutions Chapter 7 ...
Page 161: ...Glossary Chapter 8 ...
Page 181: ...Notes Piranha QABrowser Reference Manual 2014 06 5 5C Notes 177 ...