9. Start the fluoroscopy. A little triangle is displayed, indicating that the Piranha has
detected the radiation.
10. The figure to the left shows the real-time
display during fluoroscopy. The image
intensifier input dose rate is measured and
the display is updated approximately every
four seconds. Tap
Hold
to "freeze" the
currently shown value in the display. The
waveform is also acquired when Hold is
activated.
11. Release hold by tapping
Hold
again.
12. Stop the fluoroscopy.
13. Read the values.
5.4.2
kVp and Dose Rate
Use the same procedure as for the image intensifier input dose rate measurement but
select
All...
parameters instead.
1. The first screen shows how the continuous updated display looks like. The little black
arrow indicates that the radiation is detected and the display is updating every four
seconds.
2. The second screen shows that the Piranha has detected that the fluoroscopy have
stopped by flashing the logo and then freeze the values. Note that the last registered
kV value may be lower than the one measured during the exposure. The last display
update may occur when the exposure is switched off and the kVp is captured on the
"falling edge". The dose rate shown after the logo flashes is the average dose rate of
the entire exposure.
Summary of Contents for Piranha
Page 1: ...Reference Manual English Version 5 5C Piranha QABrowser RTI article number 9629050 10 ...
Page 8: ...Introduction Chapter 1 ...
Page 12: ...Description of the Piranha Chapter 2 ...
Page 40: ...Description of the QABrowser Chapter 3 ...
Page 77: ...Measurement Principles Theory Chapter 4 ...
Page 91: ...Measurements with the Piranha System Chapter 5 ...
Page 151: ...Optional Accessories Chapter 6 ...
Page 155: ...Problems and Solutions Chapter 7 ...
Page 161: ...Glossary Chapter 8 ...
Page 181: ...Notes Piranha QABrowser Reference Manual 2014 06 5 5C Notes 177 ...