5. Measurements with the Piranha System
Tube Current Probes
Piranha & QABrowser Reference Manual
2014-06/5.5C
142
7. The figure to the left shows the result from
an exposure with 80 kV, 100 ms, and
25 mA. The exposure time is measured
with the MAS-2 probe.
If you would get a message as shown in the
figure to the left, the mAs probe is probably
connected in the wrong direction. Change the
direction of the mAs probe, press the reset
button on the mAs probe, tap
Reset
,
and
make a new exposure.
Since the MAS-2 probe is based on measurement of magnetic flux, false triggering may
occur due to electrical noise when the pre-heat is started and the anode starts to rotate.
If you get incorrect or inconsistent results try the following:
First start anode rotation without firing the exposure.
While the anode is rotating tap
Reset
or press corresponding button. Make the
exposure when the reset procedure is finished.
8. Tap
Wave
to view that mA waveform.
The figure to the left shows the mA
waveform. You can use the cursor to analyse
the waveform and read the mA value as well
as estimate the exposure time.
Summary of Contents for Piranha
Page 1: ...Reference Manual English Version 5 5C Piranha QABrowser RTI article number 9629050 10 ...
Page 8: ...Introduction Chapter 1 ...
Page 12: ...Description of the Piranha Chapter 2 ...
Page 40: ...Description of the QABrowser Chapter 3 ...
Page 77: ...Measurement Principles Theory Chapter 4 ...
Page 91: ...Measurements with the Piranha System Chapter 5 ...
Page 151: ...Optional Accessories Chapter 6 ...
Page 155: ...Problems and Solutions Chapter 7 ...
Page 161: ...Glossary Chapter 8 ...
Page 181: ...Notes Piranha QABrowser Reference Manual 2014 06 5 5C Notes 177 ...