102
5.3.1
kVp, Time, Dose, and Dose Rate
Use the same procedure as for the normal radiography measurement but select the
All...
parameter.
If only
Tube voltage
is selected:
5.3.2
Pulse Measurements with Piranha Dose Probe
Use the same procedure as for the normal radiography measurement. Note that if it is
difficult to get a good pulse rate reading, you may use a manual pulse rate setting, as
described under
Settings
to get a dose per pulse reading.