2. Description of the Piranha
Indicators and Connectors
Piranha & QABrowser Reference Manual
2014-06/5.5C
12
Setting Up the Piranha for the First Time
2.2
Before you use your Piranha for the first time, please do the following:
Attach the external power supply.
Charge the system for 16 hours.
Then continue according to the following section.
Setting Up the Piranha
2.3
To set up the Piranha:
1. Power on the Piranha using the power switch. Optionally you may connect the power
supply.
2. Place the Piranha under the tube or mount the holder and HVL stand for positioning
of the Piranha in the X-ray field. The stand allows you to position the Piranha (or the
external Dose Probe) and HVL filters in any angle including upside-down. Use the
light-field or other help to position the Piranha in the X-ray field. The Piranha detector
is not sensitive for different field sizes as long as the entire sensitive detector area is
irradiated, but try to keep the field size down to minimize scattering.
It is also recommended to position the Piranha in such a way that the
detector area is orientated perpendicular to the anode/cathode axis, to
avoid the heel effect.
Recommended field size is 20×40 mm.
3. Connect with Handheld via Bluetooth, or with Ocean via included USB cable or via
Bluetooth.
Summary of Contents for Piranha
Page 1: ...Reference Manual English Version 5 5C Piranha QABrowser RTI article number 9629050 10 ...
Page 8: ...Introduction Chapter 1 ...
Page 12: ...Description of the Piranha Chapter 2 ...
Page 40: ...Description of the QABrowser Chapter 3 ...
Page 77: ...Measurement Principles Theory Chapter 4 ...
Page 91: ...Measurements with the Piranha System Chapter 5 ...
Page 151: ...Optional Accessories Chapter 6 ...
Page 155: ...Problems and Solutions Chapter 7 ...
Page 161: ...Glossary Chapter 8 ...
Page 181: ...Notes Piranha QABrowser Reference Manual 2014 06 5 5C Notes 177 ...