Reference Manual
00809-0100-4024, Rev BA
September 2005
Rosemount 5600 Series
2-39
Setting the Hold Off
Distance
To set a new Hold Off distance do the following:
1. Start the configuration software.
2. In the Hold Off/New input field type the desired Hold Off distance.
Use the following formula in order to calculate the appropriate Hold Off
(H) distance:
H=1.2 L
ext
(H=0.03 L
ext
)
where:
•
L
ext
is the length of the extended cone antenna
Figure 2-57. Extended cone
antenna
CONE_EXTENSION_1B
Hold Off (H)
L
ext
1.2 in.
(30 mm)