77
Reference Manual
00809-0200-4840, Rev BA
December 2014
Section 4: Configuration
4.3.37
Poll address (D951)
This indicates the transmitter polling address for the HART protocol.
The poll address range is 0 to 15. When it is 0, the transmitter is in 4–20 mA mode. For all other
addresses, the transmitter is in multi-drop mode and the current output is fixed to 4 mA.
Field Communicator or AMS
To view or change the poll address:
1.
From the
Home
screen, select
2: Configure
.
2.
Select
2: Manual Setup
.
3.
Select
4: HART/Identity
.
4.
Select
2: Communication
.
5.
Select
1: Poll Address D951
.
6.
Input a new setting, and then select “
ENTER
” to save it.
7.
Press “
SEND
” to update the transmitter.
Rosemount 3490 Series Control Unit
To view or change the poll address:
1.
From the
Main Menu
screen, select
SETUP
.
2.
Select the transmitter (e.g. “
Tx1: 3107
”).
3.
Select
SYSTEM
.
4.
Select
FIXED
.
5.
Select
HART
.
6.
Select
Poll Address
.
7.
Follow the on-screen instructions to edit and save the new setting.
(Press the
Enter
(
) key if prompted to change the mode to “off-line”).
8.
When finished, select “
Quit
” to exit to the previous menu.
Fast Keys
2, 2, 4, 2, 1
Poll Address D951
0
0
ESC
DEL
ENTER
HELP
w
s
q
a
z
Lock
Shift
@ &
á ü
x
e
d
c
r
f
v
t
g
b
y
h
n
u
j
m
*
-
+
/
.
0
7
4
1
8
5
2
9
6
3
i
k
o
l
FN
(Field Communicator Screen)
Poll Address P951
0
Esc=Quit =Edit
(
Rosemount 3491 Screen
)
Summary of Contents for 3107
Page 2: ......
Page 4: ...Reference Manual 00809 0200 4840 Rev BA Title Page December 2014 ...
Page 16: ...6 Reference Manual 00809 0200 4840 Rev BA Section 2 Transmitter Overview December 2014 ...
Page 145: ......
Page 146: ......
Page 147: ......