95
Reference Manual
00809-0200-4840, Rev BA
Section 5: Service and Troubleshooting
December 2014
Section 5: Service and Troubleshooting
Field Communicator or AMS
To view the temperature used for calculating the
speed of sound:
1.
From the
Home
screen, select 3: Service Tools.
2.
Select 1: Alerts.
3.
Select 2: Diagnostic Data.
4.
Select 2: Speed of Sound Data.
5.
Select 2: Temperature for Calculation D915.
6.
When finished, press EXIT to exit to the previous menu.
Note
This is also available at the Fast Key sequence 2, 2, 8, 1, 6
(see
Appendix D: Field Communicator
for a full menu map).
Rosemount 3490 Series Control Unit
To view the temperature used for calculating the
speed of sound:
1.
From the
Main Menu
screen, select MONITOR.
2.
Select the transmitter (e.g. “
Tx1: 3107
”).
3.
Select DIAGNOSTICS.
4.
Select Temp SoS Calc.
5.
When finished, press the ESC key to exit to the previous menu.
Note
Units are in
C or
F depending on the base units selected (see
).
Temperature for Calculation D915
67.901 degF
EXIT
HELP
(Field Communicator Screen)
Temp SoS Calc D915
67.9
F
(Rosemount 3491 Screen)
Summary of Contents for 3107
Page 2: ......
Page 4: ...Reference Manual 00809 0200 4840 Rev BA Title Page December 2014 ...
Page 16: ...6 Reference Manual 00809 0200 4840 Rev BA Section 2 Transmitter Overview December 2014 ...
Page 145: ......
Page 146: ......
Page 147: ......