71
Reference Manual
00809-0200-4840, Rev BA
December 2014
Section 4: Configuration
4.3.31
Target pulses (P044)
Target Pulses
(
P044
) is the number of ultrasonic pulses in each burst sent from the transmitter
every second.
The factory default “
Automatic
” setting allows the transmitter to decide the number of pulses.
Alternatively, select a number in the range 4 to 32.
Note
See
Pulses in Use
(
) for the actual number of pulses used.
Field Communicator or AMS
To view or change the number of pulses in a burst:
1.
From the
Home
screen, select
2: Configure
.
2.
Select
2: Manual Setup
.
3.
Select
7: Advanced Processing
.
4.
Select
5: Set Target Pulses
.
5.
Select
1: Target Pulses P044
.
6.
Select a new setting, and then select “
ENTER
” to save it.
7.
Press “
SEND
” to update the transmitter.
Rosemount 3490 Series Control Unit
To view or change the number of pulses in a burst:
1.
From the
Main Menu
screen, select
SETUP
.
2.
Select the transmitter (e.g. “
Tx1: 3107
”).
3.
Select
ENGINEERING
.
4.
Select
ADVANCED
.
5.
Select
Target Pulses
.
6.
Follow the on-screen instructions to edit and save the new setting.
(Press the
Enter
(
) key if prompted to change the mode to “off-line”).
7.
Select “
Quit
” to exit to the previous menu.
Fast Keys
2, 2, 7, 5, 1
Target Pulses P044
AUTOMATIC
27
28
29
30
31
32
AUTOMATIC
AUTOMATIC
ENTER
ESC
(Field Communicator Screen)
Target Pulses P043
Automatic
Esc=Quit =Edit
(Rosemount 3491 Screen)