6-12
October 1999 Rosemount Analytical 748250-L
Model 880A Non-Dispersive Infrared Analyzer
function which communicates with the computer via an 8-bit buss arrangement. This
option consists of two form C contacts rated 3A-125/250 VAC or 5A-30 VDC
(resistive). This circuit board satisfies a dual alarm requirement, as it provides two
medium power relays that can be independently controlled from the central processor.
Also, the board can be used to connect user-supplied solenoid valves to zero and
span calibration gases for one-man calibration. Provision is made to assign a specific
address in the range 0 through 7 using jumpers.
6.8.11 I
SOLATED
R
EMOTE
I
NPUT
/O
UTPUT
B
OARD
(DWG 624251)
The remote range change board is a peripheral circuit function which communicates
with the computer via an 8-bit buss arrangement. This assembly provides isolated
two-way communication between the host instrument and external user devices.
Provision is made to assign a specific address in the range 0 through 7 using jumpers.
6.8.12 A
UTO
ZERO/SPAN B
OARD
(DWG 624599)
The auto ZERO/SPAN board is a peripheral circuit function which communicates with
the computer via an 8-bit bus. With the appropriate software it satisfies the auto
ZERO/SPAN requirement. The assembly provides six form C relay contact outputs,
four of which are suitable for medium power requirements, the remaining two are
relegated to alarm or indicator functions. Snubbers are provided for the medium power
relays. Provision is made to assign a specific address in the range 0 through 7 using
jumpers.
6.8.13 C
URRENT
O
UTPUT
B
OARD
(DWG 624092)
This board changes the instrument voltage output to an isolated current output for use
with external recorders or data gathering systems.
6.8.14 A
UTO
R
ANGE
C
HANGE
This option couples the Isolated Remote Input/Output Board and the Alarm Board into
a single function. When the alarm setpoint is triggered, the remote I/O board will
automatically change ranges according to preset configurations.
JUMPER CONFIGURATION
OPTION BOARD
ADDRESS
FUNCTION
Dual Alarm
E1 + E2
E5-E7, E9-E10
Dual Alarm - Fail Safe
E1 + E2
E6-E7, E8-E10
Calibration Gas Control
E1 + E4
E5-E7, E9-E10
Auto Zero/Span
E1
--------------------
Isolated Remote I/O
E4
--------------------
T
ABLE
6-2. J
UMPER
C
ONFIGURATIONS FOR
O
PTION
B
OARDS