I
NITIAL
S
TARTUP AND
C
ALIBRATION
748250-L Rosemount Analytical October 1999
3-23
Model 880A Non-Dispersive Infrared Analyzer
T
ABLE
3-1. R
EMOTE
R
ANGE
I/O D
ESIGNATION
T
ABLE
3-2. R
EMOTE
R
ANGE
I/O B
INARY AND
D
ECIMAL
C
ODING
OUTPUT
INPUT
J2
PIN
#
DESIGNATION
J3
PIN #
DESIGNATION
1
Common
1
Range selection in remote
2
Range I.D.
2
Range selection in remote
3
Range I.D.
3
Range selection in remote
4
Not Used
4
Not Used
5
Not Used
5
Not Used
6
Not Used
6
Not Used
7
Not Used
7
Auto-Cal request
8
AUTO-CAL status
8
Not Used
9
REMOTE/LOCAL status
9
Common
Note: The Auto-Cal request input is level triggered, therefore, it is the responsibility
of the user to verify that the input is brought low before the analyzer
completes the Auto-Cal process.
MODE
RANGE
J3 PIN 3
J3 PIN 2
J3 PIN 1
BIN
R3
0
1
1
BIN
R2
0
1
0
BIN
R1
0
0
1
DEC
R3
1
0
0
DEC
R2
0
1
0
DEC
R1
0
0
1
1 = High Voltage Pulse (5 to 24 VDC)
0 = Low