I
NITIAL
S
TARTUP AND
C
ALIBRATION
3-4
October 1999 Rosemount Analytical 748250-L
Model 880A Non-Dispersive Infrared Analyzer
The SHIFT key in conjunction with the ENTER key will return to Run
Mode from any function screen except: 1) the [CALCULATING
ZERO/SPAN] screen and 2) during an auto calibration cycle.
SHIFT/ENTER during operation of zero and span functions will turn off the appropriate
solenoid valve, if connected, for instruments with the Calibration Gas Control or Auto
Zero/Span.
The SHIFT/ENTER combination is the Escape feature.
3.3.3 U
SER
-P
ROGRAMMABLE
K
EYS
Refer to Figure 3-2. F1, F2, SHIFT/F1 and SHIFT/F2 are software-programmable keys
which can be user-programmed to access any frequently used display or sub-menu
for the following modes: Range, Diagnostics, Linearization, Auto Zero/Span, Remote
Range I/O or Alarm, provided the option board selected is still present.
To use this feature, the function keys must be preprogrammed by the user through the
following steps:
1. Access a display or sub-menu that will be frequently used by following the steps in
the particular set of instructions given in the figures in this section until the desired
display is obtained.
2. Press F1, F2, SHIFT/F1 or SHIFT/F2 to program the analyzer to return to this
display from the RUN mode. This will assign F1, F2, SHIFT/F1 or SHIFT/F2 to this
particular display, and will retain those assignments until the key or combination of
keys is reprogrammed using the same procedure described in this section. The
analyzer acknowledges this command by flashing [**KEY SAVED**] on the display.
3. Exit to the RUN mode display by completing the remaining steps in the figure
chosen in Step 1.
4. When the analyzer returns to the RUN mode display, press the key or keys
selected in Step 2 (F1, F2, SHIFT/F1 or SHIFT/F2) to check the setup. The
analyzer will return to the display or sub-menu selected in Step 1.
5. Press SHIFT/ENTER to return to the RUN mode.
Note:
The programmable keys cannot be assigned to the zero or span screens since
these screens are already single-key accessed by the ZERO and SPAN keys,
respectively, on the front panel.
To reprogram the key or keys selected in Step 2, repeat Steps 1 through 5 for another
display or sub-menu.
For example, if the GAIN is frequently changed, access the RANGE sub-menu to
access the GAIN display and press the F1 key. Press SHIFT/ENTER to return to the
SHIFT
ENTER
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com