748250-L Rosemount Analytical October 1999
3-17
Model 880A Non-Dispersive Infrared Analyzer
Alarm:
MN (Manual) or
AT (Automatic)
Alarm 1 (15 sec.)
{Followed by]
Alarm 2 (15 sec.)
Alarm ON
or OFF
ALARM 1 MN OFF
ALARM 1 and ALARM 2 can be toggled with the up and down arrows to either AT
(automatic) or MAN (manual). In the AUTO (automatic) setting, an alarm relay will be
activated should an alarm condition occur. Alarms are calculated in the AUTO mode
on the basis of parameter settings. The MANUAL mode is the test mode and alarms
are not scanned by the CPU. In the MANUAL (test) mode, the ALARM ON/OFF can
be toggled with the up and down arrows to set and reset the alarm latch.
The Fail-Safe not only sets the alarm when an alarm condition is present, but also in
the event of a power outage.
3.11.1 STATUS D
ISPLAY
Refer to Figure 3-12. The STATUS display shows the alarms, error messages and
security lockout status. (See explanations of error messages in Section 5.1.) The
STATUS display can be used to check the following alarm setpoints without entering
one of the MODE functions: HIGH/LOW, AUTO/MANUAL and ON/OFF.
The order of priority for error messages, security status and alarms is as follows:
[SECURITY ENABLED/DISABLED]
[ERL ]
[EL-LIN.COEFF ERR]
[E0-ZERO POT LMTS]
[E1-SPAN #1 LMTS]
[E2-SPAN #2 LMTS]
[E3-SPAN #3 LMTS]
[E4-ADC SATURATED]
[E5-ZERO DRIFT]
[E6-SPAN DRIFT]
[E7-GAIN TOO HIGH]
[RMT: R/L]
[ALARM 1 AUTO/{MAN ON/OFF}]
[ALARM 2 AUTO/{MAN ON/OFF}]
[MAN-CAL PRESENT]
[AUTOCAL: ON/OFF]
[CURRENT 0/4 SP ON/OFF]
F
IGURE
3-12. S
TATUS
D
ISPLAY