2.10 OPTION BOARDS
The following option boards may be ordered factory installed or as kits from the factory
at a later date: Alarm, Current Output, Calibration Gas Control, Auto Zero/Span and
Remote Range I/O. The boards are supplied with outlet connector mating plugs
installed on each board for field wiring. Attach the cable (customer supplied) to the
outlet connector plug per the appropriate schematic diagram for each option board.
If an option board has been ordered installed at the factory, this board will be inserted
into one of
five slots inside the rear of the analyzer (See Figure 2-4). Each option will require a
cable (user-provided) which connects to a female plug. The female plug, in original
packaging, is attached to the appropriate terminal block on the option board. If the
instrument came equipped with one option, the interconnect cable will be in place for
all options. The Alarm, Auto Zero/Span, Calibration Gas Control and Remote Range
Change Boards have jumper-selectable addresses (Figures 2-5, 2-7 and 2-8, Section
6.8 and Table 6-2).
TB1
GND
NEUT/L2
HOT/L1
+ - GND
TB2
F1 4A
RESETTABLE
CIRCUIT
BREAKER
PURGE GAS
OUTLET
REFERENCE
OUT IN
SAMPLE
OUT IN
OPTIONS CONNECTIONS
RECORDER
OUTPUT
POWER
PURGE GAS
IN
OPTION BOARDS INSTALLED WITH COMPONENT SIDE TO THE LEFT.
Auto Zero/Span Control Board
Isolated V/I Board
Dual Alarm Board
Isolated Remote Control Board
Calibration Gas Control Board
Recorder Output
Hook-Up
Power
Hook-UP