R
OUTINE
S
ERVICING
6-2
October 1999 Rosemount Analytical 748250-L
Model 880A Non-Dispersive Infrared Analyzer
10.
Remove any contaminants from optical filters with a lint free cloth soaked in
acetone. Do not use alcohol or other solvents. Allow to air dry.
11. To install the cells, fit into position. Make sure that the o-rings seat properly. Move
the motor/source assembly back into position. Make sure that the o-rings seat
properly.
12. Replace the two screws which hold the motor/source assembly to the optical bench
plate. Do not overtighten.
13. Replace the two Motor/Source Assembly support brackets.
14. Install the sample lines.
15. Check for leaks as instructed in Section 2.8. Take corrective action if necessary.
16. Replace desiccant if required (see Section 6.2).
17. Perform source balance and source alignment adjustments per Sections 5.6 and
5.7.
6.1.2 S
HORT
C
ELL
C
ONFIGURATIONS
Refer to Figure 6-1B.
1. Slide chassis out.
2. Remove sample lines from the sample cell assembly.
3. Remove the two screws holding the motor/source assembly to the optical bench
plate.
4. Remove the two screws holding the end cap/optical filter assembly and sample cell
assembly to the detector. Gently move the motor/source assembly away from the
detector. The sample cell assembly, end cap/optical filter assembly and o-rings will
now be free.
5. Rinse the cell with acetone. If this does not remove all foreign matter, use a soft
brush. Do not use any metallic object inside the cell.
6. After all matter has been removed, rinse the cell with distilled water and allow to air
dry. Do not use towels.
7. Inspect the cell inside by holding it up to a bright light. If particles are seen, repeat
Steps 5 and 6 as often as necessary.
8. After cleaning cell, examine o-rings between the detector and sample cell
assembly, and between the sample cell assembly and end cap/optical filters
assembly, and the compression gaskets between the end cap/optical filters
assembly and plate assembly, replace If damaged.
9. Remove any contaminants from windows with a lint free cloth soaked in acetone.
Do not use alcohol or other solvents. Allow to air dry.
10. To install the cells, replace the two retaining screws holding the end cap/optical
filter assembly and sample cell assembly to the detector. Verify that the o-rings