P
REFACE
P-4
October 1999 Rosemount Analytical 748250-L
Model 880A Non-Dispersive Infrared Analyzer
O
PTIONS
S
PECIFICATIONS
A
LARM
Alarm Relay Outputs:
Two single point, field programmable high or low, deadband 5%
to 20% of fullscale.
Two form C contacts rated 3A-125/250 VAC or 5A-30 VDC
(resistive)
Z
ERO
/S
PAN
Relay Outputs:
Two front panel actuated contact closures
Two form C contacts rated 3A-125/250 VAC or 5A-30 VDC
(resistive)
A
UTO
Z
ERO
/S
PAN
Relay Outputs:
Four form C contacts rated 3A-125/250 VAC or 5A-30 VDC
(resistive), field programmable frequency and duration of
closure.
Two form A contact closures for indication of insufficient zero
and span adjustment, rated (resistive load):
Max switching power: 10 Watts
Max switching voltage: 30 VDC
Max switching current: 0.5 A
R
EMOTE
R
ANGE
C
HANGE
& ID
Three remotely changeable ranges with positive identification.
Binary code field selectable.
R
ELAY
O
UTPUTS
:
Eight form A contacts rated (resistive load):
Max switching power: 10 Watts
Max switching voltage: 30 VDC
Max switching current: 0.5 A
I
NPUTS
:
Eight optical couplers
Input Range: +5 VDC to +24 VDC