6.5 CHOPPER MOTOR ASSEMBLY (P/N 652605)
To remove the chopper motor assembly, do the following :
6.5.1 L
ONG
C
ELL
C
ONFIGURATIONS
1. Refer to Figure 6-1A. Remove the two screws holding the motor/source assembly
(with the end cap assembly attached) to the optical bench base plate. Support the
cells and gently slide the motor/source assembly away from the cells.
2. Remove the two screws holding the end cap assembly to the motor/source
assembly.
3. Refer to Figure 6-2. Remove chopper cover.
4. Remove chopper blade.
5. Remove two screws from rear of motor and remove motor.
6.5.2 S
HORT
C
ELL
C
ONFIGURATIONS
1. Refer to Figure 6-1B. Remove the two screws holding the motor/source assembly
(with the plate assembly attached) to the optical bench base plate. Gently slide the
motor/source assembly away from the cells.
2. Remove the two screws holding the plate assembly to the motor/source assembly.
3. Refer to Figure 6-2. Remove chopper cover.
4. Remove chopper blade.
5. Remove two screws from rear of motor and remove motor.
F
IGURE
6-2. M
OTOR
/S
OURCE
A
SSEMBLY
Chopper Motor
652605
O-Ring
900406
Jam Nut
801682
Washer, Nylon
901077
Screw
901753
Shutter Adjustment Screw
901351
Chopper Cover
624167
Chopper Blade
620300
Chopper Housing
624166 (Sapphire Windows)
624411 (Irtran Windows)
Teflon
Spacers
Source Housing
622706
Source (Matched Pair)
624442