
2
U
NPACKING AND
I
NSTALLATION
748250-L Rosemount Analytical October 1999
2-1
Model 880A Non-Dispersive Infrared Analyzer
2.1 UNPACKING
Carefully examine the shipping carton and contents for signs of damage. Immediately
notify the shipping carrier if the carton or its contents are damaged. Retain the carton
and packing material until the instrument is operational
2.2 LOCATION
Locate the analyzer in a weather-protected, non-hazardous location free from
vibration. For best results mount the analyzer near the sample stream to minimize
sample-transport time. Refer to Installation Drawing 624190.
If equipped with P/N 624446 air purge kit and installed with user-provided
components per Instructions 015-748157, the analyzer may be located in a Class I,
Division 2 area as defined by the National Electrical Code (ANSI/NFPA 70). This kit is
designed to provide Type Z protection in accordance with Standard ANSI/NFPA
496-1989, Chapter 2, when sampling nonflammable gases. For flammable samples,
the instrument must be equipped with a continuous dilution purge system in
accordance with ANSI/NFPA 496-1989, Chapter 8.
2.3 VOLTAGE REQUIREMENTS
For safety and proper performance, this instrument must be connected to a
properly grounded three-wire source of electrical power.
This instrument was shipped from the factory configured to operate on 115 VAC,
50/60 Hz electric power. For operation on 230 VAC, 50/60 Hz, voltage select switches
S1, S2, located on Power Supply Board (Figure 2-1) and, if installed, S3, located on
the Case Heater Temperature Control Board Figure 2-2 (see Figure 7-2, Case Heater
Temperature Control Assembly) must be in the 230 VAC position. Power consumption
is 350 watts.
WARNING: ELECTRICAL SHOCK HAZARD
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com