I
NITIAL
S
TARTUP AND
C
ALIBRATION
748250-L Rosemount Analytical October 1999
3-9
Model 880A Non-Dispersive Infrared Analyzer
Cal Gas (Span)
Concentration in ppm or
%/Volume.
Zero Gas Offset as a %
of fullscale.
Time Constant: 0.5 to
20 seconds
SHIFT
CAL GAS = ppm
SHIFT
ZR. OFFSET: 0.00
SHIFT
TIME CONST = XX
RANGE=X CMP SO2
Digital Gain
(X1, X2, X4, X8)
ENTER
GAIN= X
SHIFT
SHIFT
FS - XX ppm
Fullscale Range in %
or ppm
RANGE
and will move cursor right/left, one character at a time. and will go to next function.
and will change (toggle) settings. and will return analyzer to RUN MODE.
ENTER
SHIFT
SHIFT
3.6.2 C
OMPONENT OF
I
NTEREST
The component of interest is designated by a 3-digit group of letters or numbers. This
gas name or designation may be selected for each range by placing the cursor under
the desired digit [NNN] and selecting a letter or number with
↑
or
↓
key. This name
will appear on the display when the analyzer is in the run mode.
3.6.3 G
AIN
In the [GAIN=X] display, an amplifier gain of 1, 2, 4, or 8 can be selected for each
range with the
↑
or
↓
key depending on the sensitivity desired (Refer to Section
5.5). Range 3 is normally the least sensitive range.
F
IGURE
3-6. R
ANGE
M
ODE
Other ranges are generally set with gains that are proportional to their relative fullscale
spans. Thus, if range 1 is 0 to 50 ppm CO and range 3 is 0 to 500 ppm CO, then the
respective gains will usually be 8 and 1.
3.6.4 R
ANGE
, F
ULLSCALE
In the [FS=XXXX ppm *] or [FS=XX.X % *] display, up to a four-digit fullscale value is
entered in ppm (parts per million).