748250-L Rosemount Analytical October 1999
6-1
Model 880A Non-Dispersive Infrared Analyzer
Servicing requires access to live parts which can cause death or serious injury.
Refer servicing to qualified personnel.
Note:
Before servicing analyzer, disconnect power and shut off sample flow to unit.
6.1 CELL REMOVAL, CLEANING AND INSTALLATION
6.1.1 L
ONG
C
ELL
C
ONFIGURATIONS
Refer to Figure 6-1A.
1. Slide chassis out.
2. Remove sample lines from the end cap assembly and the end cap/optical filter
assembly.
3. If installed, remove the two motor/source assembly support brackets.
4. Remove the two screws holding the motor/source assembly to the optical bench
plate.
5. Support the cells and gently move the motor/source assembly away from the
detector. The cells and its o-rings will now be free.
6. Rinse the cell with acetone. If this does not remove all foreign matter, use a soft
brush. Do not use any metallic object inside the cell because it will scratch the gold
plating. Loss of gold plating may require cell replacement.
7. After all matter has been removed, rinse the cell with distilled water and allow to air
dry. Do not use towels.
8. Inspect the cell inside by holding it up to a bright light. If particles are seen, repeat
Steps 6 and 7 as often as necessary.
9. After cleaning cells, examine o-rings at the end cap assembly and end cap/optical
filter assembly and replace If damaged.
WARNING: ELECTRICAL SHOCK HAZARD