R
OUTINE
S
ERVICING
6-6
October 1999 Rosemount Analytical 748250-L
Model 880A Non-Dispersive Infrared Analyzer
6.6. MICRO BOARD REPLACEMENT
All calibration constants and settings stored in non-volatile memory must be changed
when the Micro Board is replaced. These procedures are given in Section 3.
Additionally, the ZERO potentiometer and the three SPAN potentiometers must be
resynchronized with software. To resynchronize the ZERO potentiometer follow the
steps listed below:
1. Follow the instructions in Section 3-8 for standard analyzer or Section 3-13 for
analyzer with Calibration Gas Control option and access the [ZR=X.X PS=XX%]
display.
2. Press SHIFT then
↓
.
3. Proceed with the ZERO calibration.
4. The SPAN potentiometer must be re- synchronized for all three ranges as follows:
5. Choose the first range by selecting a range using the range parameters in Figure
3-6.
6. Follow the instructions in Section 3-9 for standard analyzer or Section 3-14 for
analyzer with Auto ZERO/SPAN option and access the [ X.XX XX % 100 ] display.
7. Press SHIFT then
↓
.
Proceed with the SPAN calibration.
6.7 DETECTOR REPLACEMENT
6.7.1 R
EMOVAL
- L
ONG
C
ELL
C
ONFIGURATIONS
Refer to Figure 6-3A.
1. Slide chassis out.
2. Remove sample lines from the end cap assembly and end cap/optical filter
assembly.
3. Remove detector cover.
4. While supporting the cells, remove the four screws holding the detector assembly
to the optical bench base plate.. The detector assembly, detector heater, detector
plate, cells and o-rings are now free
5. Disconnect ribbon cables.
6. Remove oscillator board.
Refer to Figure 6-3C.