ENTER
Toggle and keys to select YES or NO.
Toggle and keys to select between 0 to 20
mA and 4 to 20 mA.
and will return analyzer to RUN MODE.
SHIFT
ENTER
SHIFT
CURRENT: X 20mA
ZR/SPN SUPR: YES
XX TO XXX FS
Note: Current Output is factory set at 4 to 20 mA. If an adjustment is needed, Zero (R2, labeled OFFSET on the board)
and Span (R1) potentiometers are located on the Current Output Board (Figure 2-6).
3.12 CURRENT OUTPUT
Refer to Figure 3-13. The Model 880A Analyzer has a standard 0 to 20 or 4 to 20 mA
current output with zero span suppression. This function can represent any
suppressed range with at least a 25% span. For example. a valid range could be 0%
to 25%, 28% to 61% or 33% to 100%. When the Zero Span Suppression is off (NO),
the analyzer defaults to the 0% to 100% range.
F
IGURE
3-13. C
URRENT
O
UTPUT
M
ODE
3.13 ZERO/SPAN CALIBRATION WITH CALIBRATION GAS
CONTROL
The Calibration Gas Control allows one-man calibration. This option consists of two
form C contacts, rated 3A-125/250 VAC or 5A-30 VDC (resistive). These contacts are
connected to solenoid valves (customer supplied) which will turn zero and span
calibration gases on and off when activated. Simply press ZERO or SPAN to open the
appropriate valve, thus allowing the gas to flow through the instrument. Press ENTER
to initiate the calibration process. At the conclusion of calibration, the valve is closed
and the instrument returns to Run Mode. See Section 3.16 for more information.
Note:
For instruments with Calibration Gas Control or Auto Zero/Span, press ENTER
to perform the selected calibration. Press SHIFT/ ENTER to abort the function,
turn off the relay for this valve, and maintain previous calibration settings.