System Overview
R&S
®
ZVT
84
Getting Started 1300.0045.62 ─ 10
Calibration unit
Recommended for Frequency range
Connector type
No. of ports
Order no.
R&S ZN-Z151
R&S ZVB4, R&S
ZVB8
100 kHz to 8.5 GHz type N (f)
2
1317.9134.72
R&S ZN-Z151
R&S ZVB4, R&S
ZVB8
100 kHz to 8.5 GHz SMA (f)
2
1317.9134.32
R&S ZN-Z153
R&S ZVB4, R&S
ZVB8
100 kHz to 8.5 GHz SMA (f)
4
1319.6178.34
The multiport calibration units R&S ZV-Z58, R&S ZV-Z59, R&S ZN-Z152 and R&S ZN-
Z154 are especially suited for R&S ZVT vector network analyzers.
Calibration unit
Recommended for Frequency range
Connector type
No. of ports
Order no.
R&S ZV-Z58
R&S ZVT8
300 kHz to 8 GHz
type N
8
1164.0638.78
R&S ZV-Z58
R&S ZVT8
300 kHz to 8 GHz
3.5 mm
8
1164.0638.78
R&S ZV-Z59
R&S ZVT20
10 MHz to 20 GHz
3.5 mm
6
1164.0450.36
R&S ZN-Z152
R&S ZVT8
100 kHz to 8.5 GHz SMA (f)
6
1319.6003.36
R&S ZN-Z153
R&S ZVT8
100 kHz to 8.5 GHz SMA (f)
4
1319.6178.34
The units contain calibration standards that are electronically switched when a calibra-
tion is performed. The calibration kit data for the internal standards are also stored in
the calibration unit, so that the analyzer can calculate the error terms and apply the cal-
ibration without any further input.
Advantages of automatic calibration
Automatic calibration is generally faster and more secure than manual calibration,
because:
●
There is no need to connect several standards manually. The number of connec-
tions to be performed quickly increases with the number of ports; see "TOSM Cali-
bration".
●
Invalid calibrations due to operator errors (e.g. wrong standards or improper con-
nections) are almost excluded.
●
No need to handle calibration kit data.
●
The internal standards don't wear out because they are switched electronically.
Calibration