System Overview
R&S
®
ZVT
78
Getting Started 1300.0045.62 ─ 10
●
S
<mout><min>11
is the mixed mode input reflection coefficient, defined as the ratio of
the wave quantities b
1
(mode mout) to a
1
(mode min), measured at PORT 1 (for-
ward measurement with matched output and a
2
= 0).
●
S
<mout><min>21
is the mixed mode forward transmission coefficient, defined as the
ratio of the wave quantities b
2
(mode mout) to a
1
(mode min) (forward measure-
ment with matched output and a
2
= 0).
●
S
<mout><min>12
is the mixed mode reverse transmission coefficient, defined as the
ratio of the wave quantities b
1
(mode mout) (reverse measurement with matched
input, b
1
' in the figure above and a
1
= 0) to a
2
(mode min).
●
S
<mout><min>22
is the mixed mode output reflection coefficient, defined as the ratio of
the wave quantities b
2
(mode mout) (reverse measurement with matched input, b
2
'
in the figure above and a
1
= 0) to a
2
(mode min), measured at PORT 2.
If <mout> is different from <min>, the S-parameters are called mode conversion fac-
tors.
3.3.2.14
Mixed Mode Parameters for Different Test Setups
Which types of mixed mode parameter are available depends on the measured device
and the port configuration of the analyzer. The following examples of mixed more
parameters can all be obtained with a 4-port analyzer.
1. DUT with only single-ended ports: No balanced port definition necessary, the ana-
lyzer provides single-ended multiport parameters.
2. DUT with one balanced port: Only reflection and mode conversion measurements
with differential and common mode parameters.
3. 3. DUT with one balanced and one single-ended port.
Measured Quantities