System Overview
R&S
®
ZVT
59
Getting Started 1300.0045.62 ─ 10
●
Smith charts are used for the display format "Smith". They show a complex quan-
tity like polar diagrams but with grid lines of constant real and imaginary part of the
impedance.
●
Inverted Smith charts are used for the display format "Inverted Smith". They show
a complex quantity like polar diagrams but with grid lines of constant real and
imaginary part of the admittance.
The analyzer allows arbitrary combinations of display formats and measured quantities
("Trace – Measure"). Nevertheless, in order to extract useful information from the data,
it is important to select a display format which is appropriate to the analysis of a partic-
ular measured quantity; see
chapter 3.2.4.6, "Measured Quantities and Display For-
3.2.4.1
Cartesian Diagrams
Cartesian diagrams are rectangular diagrams used to display a scalar quantity as a
function of the stimulus variable (frequency / power / time).
●
The stimulus variable appears on the horizontal axis (x-axis), scaled linearly
(sweep types "Lin Frequency", "Power", "Time", "CW Mode") or logarithmically
(sweep type "Log Frequency").
●
The measured data (response values) appears on the vertical axis (y-axis). The
scale of the y-axis is linear with equidistant grid lines although the y-axis values
may be obtained from the measured data by non-linear conversions.
The following examples show the same trace in Cartesian diagrams with linear and
logarithmic x-axis scaling.
Screen Elements