Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
122
User Manual 1179.3315.02 ─ 01
TRIGger[:SEQuence]:SOURce
<Source>
This command selects the trigger source.
Note that the availability of trigger sources depends on the connected instrument.
Note on external triggers:
If a measurement is configured to wait for an external trigger signal in a remote control
program, remote control is blocked until the trigger is received and the program can
continue. Make sure this situation is avoided in your remote control programs.
Parameters:
<Source>
IMMediate
Free Run
EXTernal
Trigger signal from the "Trigger Input" connector.
MAGNitude
For (offline) input from a file, rather than an instrument.
The trigger level is specified by
MAIT
For trigger information stored as markers in an
.iqx
file.
MANual
Only available for a connected R&S
RTP:
Any trigger settings in the R&S
VSE software are ignored; only
trigger settings defined on the connected instrument are consid-
ered. Thus, you can use the more complex trigger settings avail-
able on an R&S
RTP.
*RST:
IMMediate
Example:
TRIG:SOUR EXT
Selects the external trigger input as source of the trigger signal
Manual operation:
See
See
See
See
See
See
See
TRIGger[:SEQuence]:OSCilloscope:COUPling
<CoupType>
Configures the coupling of the external trigger to the oscilloscope.
Parameters:
<CoupType>
Coupling type
DC
Direct connection with 50
Ω
termination, passes both DC and
AC components of the trigger signal.
Configuring the Measurement