Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
121
User Manual 1179.3315.02 ─ 01
Parameters:
<TriggerLevel>
Range:
-130 dBm to 30 dBm
*RST:
-20 dBm
Default unit: DBM
Example:
TRIG:LEV:IQP -30DBM
TRIGger[:SEQuence]:LEVel:RFPower
<TriggerLevel>
This command defines the power level the RF input must exceed to cause a trigger
event. Note that any RF attenuation or preamplification is considered when the trigger
level is analyzed. If defined, a reference level offset is also considered.
The input signal must be between 500
MHz and 8
GHz.
Parameters:
<TriggerLevel>
For details on available trigger levels and trigger bandwidths see
the data sheet.
*RST:
-20 dBm
Default unit: DBM
Example:
TRIG:LEV:RFP -30dBm
TRIGger[:SEQuence]:RFPower:HOLDoff
<Time>
This command defines the holding time before the next trigger event. Note that this
command is available for any trigger source, not just RF Power.
Note that this command is maintained for compatibility reasons only. Use the
TRIGger[:SEQuence]:IFPower:HOLDoff
on page 119 command for new remote
control programs.
Parameters:
<Time>
Default unit: S
TRIGger[:SEQuence]:SLOPe
<Type>
Parameters:
<Type>
POSitive | NEGative
POSitive
Triggers when the signal rises to the trigger level (rising edge).
NEGative
Triggers when the signal drops to the trigger level (falling edge).
*RST:
POSitive
Example:
TRIG:SLOP NEG
Manual operation:
See
Configuring the Measurement