OUTPut<up>:TRIGger<tp>:LEVel
<Level>
This command defines the level of the (TTL compatible) signal generated at the trigger
output.
This command works only if you have selected a user defined output with
.
Suffix:
<up>
.
1..n
<tp>
Selects the trigger port to which the output is sent.
Parameters:
<Level>
HIGH
5 V
LOW
0 V
*RST:
LOW
Example:
OUTP:TRIG2:LEV HIGH
OUTPut<up>:TRIGger<tp>:OTYPe
<OutputType>
This command selects the type of signal generated at the trigger output.
Suffix:
<up>
.
1..n
<tp>
Selects the trigger port to which the output is sent.
Parameters:
<OutputType>
DEVice
Sends a trigger signal when the R&S
VSE has triggered inter-
nally.
TARMed
Sends a trigger signal when the trigger is armed and ready for
an external trigger event.
UDEFined
Sends a user defined trigger signal. For more information see
*RST:
DEVice
OUTPut<up>:TRIGger<tp>:PULSe:IMMediate
This command generates a pulse at the trigger output.
Suffix:
<up>
.
Selects the trigger port to which the output is sent.
<tp>
1..n