Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
98
User Manual 1179.3315.02 ─ 01
[SENSe:]MIXer<x>:FREQuency:HANDover
<Frequency>
This command defines the frequency at which the mixer switches from one range to
the next (if two different ranges are selected). The handover frequency for each band
can be selected freely within the overlapping frequency range.
This command is only available if the external mixer is active (see
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<Frequency>
Default unit: HZ
Example:
MIX ON
Activates the external mixer.
MIX:FREQ:HAND 78.0299GHz
Sets the handover frequency to 78.0299 GHz.
[SENSe:]MIXer<x>:FREQuency:STARt
This command sets or queries the frequency at which the external mixer band starts.
Suffix:
<x>
.
1..n
irrelevant
Example:
MIX:FREQ:STAR?
Queries the start frequency of the band.
[SENSe:]MIXer<x>:FREQuency:STOP
This command sets or queries the frequency at which the external mixer band stops.
Suffix:
<x>
.
1..n
irrelevant
Example:
MIX:FREQ:STOP?
Queries the stop frequency of the band.
[SENSe:]MIXer<x>:HARMonic:BAND:PRESet
This command restores the preset frequency ranges for the selected standard wave-
guide band.
Note:
Changes to the band and mixer settings are maintained even after using the
[PRESET] function. Use this command to restore the predefined band ranges.
Configuring the Measurement