Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
96
User Manual 1179.3315.02 ─ 01
Example:
SYST:COMM:RDEV:OSC:VFIR?
Usage:
Query only
Using External Mixers
The commands required to work with external mixers in a remote environment are
described here. Note that these commands require the connected instrument to have
an external mixer option installed and an external mixer to be connected to the connec-
ted instrument.
Programming Example: Working with an External Mixer
...................................... 103
Basic Settings
The basic settings concern general usage of an external mixer.
...............................................................................................96
..........................................................................................96
.........................................................................................97
............................................................................................ 97
[SENSe:]MIXer<x>[:STATe]
<State>
Activates or deactivates the use of a connected external mixer as input for the mea-
surement. This command is only available if the optional External Mixer is installed and
an external mixer is connected.
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
MIX ON
[SENSe:]MIXer<x>:BIAS:HIGH
<BiasSetting>
This command defines the bias current for the high (last) range.
This command is only available if the external mixer is active (see
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<BiasSetting>
*RST:
0.0 A
Default unit: A
Configuring the Measurement