Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
97
User Manual 1179.3315.02 ─ 01
[SENSe:]MIXer<x>:BIAS[:LOW]
<BiasSetting>
This command defines the bias current for the low (first) range.
This command is only available if the external mixer is active (see
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<BiasSetting>
*RST:
0.0 A
Default unit: A
[SENSe:]MIXer<x>:LOPower
<Level>
This command specifies the LO level of the external mixer's LO port.
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<Level>
numeric value
Range:
13.0 dBm to 17.0 dBm
Increment: 0.1 dB
*RST:
15.5 dBm
Example:
MIX:LOP 16.0dBm
Mixer Settings
The following commands are required to configure the band and specific mixer set-
tings.
[SENSe:]MIXer<x>:FREQuency:HANDover
........................................................................ 98
[SENSe:]MIXer<x>:FREQuency:STARt
.............................................................................. 98
[SENSe:]MIXer<x>:FREQuency:STOP
...............................................................................98
[SENSe:]MIXer<x>:HARMonic:BAND:PRESet
.................................................................... 98
[SENSe:]MIXer<x>:HARMonic:BAND
.................................................................................99
[SENSe:]MIXer<x>:HARMonic:HIGH:STATe
....................................................................... 99
[SENSe:]MIXer<x>:HARMonic:HIGH[:VALue]
................................................................... 100
[SENSe:]MIXer<x>:HARMonic:TYPE
............................................................................... 100
[SENSe:]MIXer<x>:HARMonic[:LOW]
.............................................................................. 100
...................................................................................... 101
[SENSe:]MIXer<x>:LOSS:TABLe:HIGH
............................................................................101
[SENSe:]MIXer<x>:LOSS:TABLe[:LOW]
...........................................................................101
..................................................................................... 102
..............................................................................................102
[SENSe:]MIXer<x>:RFOVerrange[:STATe]
........................................................................ 102
Configuring the Measurement