Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
100
User Manual 1179.3315.02 ─ 01
Suffix:
<x>
.
1..n
Parameters:
<State>
ON | OFF
*RST:
ON
Example:
MIX:HARM:HIGH:STAT ON
[SENSe:]MIXer<x>:HARMonic:HIGH[:VALue]
<HarmOrder>
This command specifies the harmonic order to be used for the high (second) range.
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<HarmOrder
numeric value
Range:
2 to 128 (USER band); for other bands: see band
definition
Example:
MIX:HARM:HIGH:STAT ON
MIX:HARM:HIGH 2
[SENSe:]MIXer<x>:HARMonic:TYPE
<OddEven>
This command specifies whether the harmonic order to be used should be odd, even,
or both.
Which harmonics are supported depends on the mixer type.
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<OddEven>
ODD | EVEN | EODD
ODD | EVEN | EODD
*RST:
EVEN
Example:
MIX:HARM:TYPE ODD
[SENSe:]MIXer<x>:HARMonic[:LOW]
<HarmOrder>
This command specifies the harmonic order to be used for the low (first) range.
Suffix:
<x>
.
1..n
irrelevant
Configuring the Measurement