Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
95
User Manual 1179.3315.02 ─ 01
SYSTem:COMMunicate:RDEVice:OSCilloscope:SRATe
<Rate>
Determines whether the 10
GHz mode (default) or 20
GHz mode of the connected
oscilloscope is used. The 20
GHZ mode achieves a higher decimation gain, but
reduces the record length by half.
Parameters:
<Rate>
10 GHz | 20 GHz
No other sample rate values are allowed.
*RST:
10 GHz
Default unit: HZ
Example:
TRAC:IQ:SRAT?
//Result: 100000000
TRAC:IQ:RLEN?
//Result: 3128
SYST:COMM:RDEV:OSC:SRAT 20GHZ
TRAC:IQ:SRAT?
//Result: 200000000
TRAC:IQ:RLEN?
//Result: 1564
Manual operation:
See
SYSTem:COMMunicate:RDEVice:OSCilloscope:VDEVice?
Queries whether the connected instrument is supported by the 2
GHz bandwidth
extension option(B2000).
Return values:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
SYST:COMM:RDEV:OSC:VDEV?
Usage:
Query only
SYSTem:COMMunicate:RDEVice:OSCilloscope:VFIRmware?
Queries whether the firmware on the connected oscilloscope is supported by the
2
GHz bandwidth extension (B2000) option.
Return values:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Configuring the Measurement