![Rohde & Schwarz VSE-K149 User Manual Download Page 89](http://html.mh-extra.com/html/rohde-and-schwarz/vse-k149/vse-k149_user-manual_1477887089.webp)
Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
89
User Manual 1179.3315.02 ─ 01
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<Mode>
NARRow
Performs a measurement by automatically applying all available
combinations of low and high pass filters consecutively. These
combinations all have a narrow bandwidth.
WIDE
Performs a measurement by automatically applying all available
bandpass filters consecutively. The bandpass filters have a wide
bandwidth.
Manual operation:
See
INPut<ip>:PRESelection[:STATe]
<State>
This command turns the preselector on and off.
Suffix:
<ip>
.
1 | 2
irrelevant
Manual operation:
See
INPut<ip>:RF:CAPMode
<CAPMode>
Determines how data from an oscilloscope is input to the R&S
VSE software.
This command is only available for connected oscilloscopes.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<CAPMode>
AUTO | IQ | WAVeform
IQ
The measured waveform is converted to I/Q data directly on the
R&S
oscilloscope (requires option K11), and input to the
R&S
VSE software as I/Q data.
WAVeform
The data is input in its original waveform format and converted
to I/Q data in the R&S
VSE software. No additional options are
required on the R&S
oscilloscope.
AUTO
Uses "I/Q" mode when possible, and "Waveform" only when
required by the application (e.g. Pulse measurement).
*RST:
IQ
Example:
INP:RF:CAPM WAV