![Rohde & Schwarz VSE-K149 User Manual Download Page 124](http://html.mh-extra.com/html/rohde-and-schwarz/vse-k149/vse-k149_user-manual_1477887124.webp)
Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
124
User Manual 1179.3315.02 ─ 01
Manual operation:
See
TRIGger[:SEQuence]:MAPower:HYSTeresis
<Hysteresis>
This command defines the trigger hysteresis for the (offline) magnitude trigger source
(used for input from a file).
Parameters:
<Hysteresis>
Range:
3 dB to 50 dB
*RST:
3 dB
Default unit: DB
Example:
TRIG:SOUR MAP
Sets the (offline) magnitude trigger source.
TRIG:MAP:HYST 10DB
Sets the hysteresis limit value.
Manual operation:
See
6.6.3.2
Configuring the Trigger Output
The following commands are required to send the trigger signal to one of the variable
"TRIGGER INPUT/OUTPUT" connectors on the connected instrument.
OUTPut<up>:TRIGger<tp>:DIRection
.............................................................................. 124
.................................................................................... 125
..................................................................................125
OUTPut<up>:TRIGger<tp>:PULSe:IMMediate
.................................................................. 125
OUTPut<up>:TRIGger<tp>:PULSe:LENGth
......................................................................126
OUTPut<up>:TRIGger<tp>:DIRection
<Direction>
This command selects the trigger direction for trigger ports that serve as an input as
well as an output.
Suffix:
<up>
.
irrelevant
<tp>
Parameters:
<Direction>
INPut | OUTPut
INPut
Port works as an input.
OUTPut
Port works as an output.
*RST:
INPut
Configuring the Measurement