Slope ← Trigger Settings
For all trigger sources except time, you can define whether triggering occurs when the
signal rises to the trigger level or falls down to it.
Remote command:
Hysteresis ← Trigger Settings
Defines the distance in dB to the trigger level that the trigger source must exceed
before a trigger event occurs. Setting a hysteresis avoids unwanted trigger events
caused by noise oscillation around the trigger level.
This setting is only available for "IF Power" or "Magnitude (Offline)" trigger sources.
Remote command:
TRIGger[:SEQuence]:IFPower:HYSTeresis
TRIGger[:SEQuence]:MAPower:HYSTeresis
Trigger Holdoff ← Trigger Settings
Defines the minimum time (in seconds) that must pass between two trigger events.
Trigger events that occur during the holdoff time are ignored.
Remote command:
TRIGger[:SEQuence]:IFPower:HOLDoff
TRIGger[:SEQuence]:MAPower:HOLDoff