Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
123
User Manual 1179.3315.02 ─ 01
CDLimit
Direct connection with 1
M
Ω
termination, passes both DC and
AC components of the trigger signal.
AC
Connection through capacitor, removes unwanted DC and very
low-frequency components.
*RST:
DC
[SENSe:]MSRA:CAPTure:OFFSet
<Offset>
This setting is only available for secondary applications in MSRA mode, not for the
MSRA primary application. It has a similar effect as the trigger offset in other measure-
ments.
Parameters:
<Offset>
This parameter defines the time offset between the capture buf-
fer start and the start of the extracted secondary application
data. The offset must be a positive value, as the secondary
application can only analyze data that is contained in the cap-
ture buffer.
Range:
0 to <Record length>
*RST:
0
Default unit: S
TRIGger[:SEQuence]:LEVel:MAPower
<TriggerLevel>
This command defines the power level that must be exceeded to cause a trigger event
for (offline) input from a file.
Parameters:
<TriggerLevel>
For details on available trigger levels and trigger bandwidths see
the data sheet.
Default unit: DBM
Example:
TRIG:LEV:MAP -30DBM
TRIGger[:SEQuence]:MAPower:HOLDoff
<Period>
This command defines the holding time before the next trigger event for (offline) input
from a file.
Parameters:
<Period>
Range:
0 s to 10 s
*RST:
0 s
Default unit: S
Example:
TRIG:SOUR MAGN
Sets an offline magnitude trigger source.
TRIG:MAP:HOLD 200 ns
Sets the holding time to 200 ns.
Configuring the Measurement