Remote Control
R&S
®
RTM20xx
284
User Manual 1317.4726.02 ─ 01
Usage:
Query only
TIMebase:POSition
<Offset>
Defines the trigger position (trigger offset) - the time interval between trigger point and
reference point to analize the signal some time before or after the trigger event.
See also:
Parameters:
<Offset>
Range:
Depends on time base setting
*RST:
0
Default unit: s
TIMebase:REFerence
<ReferencePoint>
Sets the reference point of the time scale (Time Reference) in % of the display. The
reference point defines which part of the waveform is shown. If the trigger position is zero,
the trigger point matches the reference point.
See also:
Parameters:
<ReferencePoint>
Range:
10 to 90
Increment: 10
*RST:
50
Default unit: %
15.2.4.3
Acquisition
.............................................................................................285
.............................................................................................287
................................................................................................287
...........................................................................................287
...............................................................................................288
AUToscale
Performs an autoset process: analyzes the enabled channel signals, and obtains appro-
priate horizontal, vertical, and trigger settings to display stable waveforms.
Usage:
Event
Asynchronous command
Command Reference