MASK:STATe
<State>
Turns the mask test mode on or off. When turning off, any temporarily stored new masks
are deleted.
Parameters:
<State>
ON | OFF
*RST:
OFF
MASK:TEST
<Test>
Starts, finishes or interrupts a mask test.
Parameters:
<Test>
RUN | STOP | PAUSe
*RST:
STOP
MASK:LOAD
<FileName>
Loads a stored mask from the specified file.
Setting parameters:
<FileName>
String parameter
Path and file name
Usage:
Setting only
MASK:SAVE
<FileName>
Saves the current mask in the specified file.
Setting parameters:
<FileName>
String parameter
Path and file name
Usage:
Setting only
MASK:SOURce
<Source>
Defines the channel to be compared with the mask.
Parameters:
<Source>
CH1 | CH2 | CH3 | CH4
CH3 and CH4 are only available on 4-channel R&S
RTM models.
*RST:
CH1
MASK:CHCopy
Creates a mask from the envelope waveform of the test source set with
.
Command Reference